OBW-220A Infant Radiant Warmer



Brief introduction:
☆Micropro cessor based servo controlled temperature system
☆Alarm functions for safety includes:
 Power failure; Temperature deviation; Temperature sensor failure; Over temperature;
☆Transparent protetor
☆Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source,the heating is quick and life-span is long.
☆Alarms are audible and visual
☆A second thermal cut-out function for more safety
☆Skin temperature sensor failure protect function avoid over stemperature

Standard configuration:
 Main body (including the radiant source, control system, infant bed, bracket),  I.V.pole, skin temperature sensor, tray, mattress, transparent protector, castors.

Technical data:
 1. Power supply: 110/220V   60/50Hz
 2.  Power input: ≤1200VA
 3. Operating condition:
    An ambient Temperature:18℃-30℃
    An ambient Relative Humidity: 30%--75%RH
    Atmospheric Pressure: 700~1060 hPa
    An ambient Air Velocity : <0.3m/s
 4. Temperature control: 32℃~38℃
 5.Temperature control accuracy: ≤0.5℃
 6.Mattress temperature uniformity : <2℃
 7.Skin temperature sensor accuracy: + 0.3℃
 8.Warm-up time: ≤30min
 9.Mattress size: 67cm× 54 cm
 10.Distance from heater to mattress: 80cm

Transport and storage:
 Environment Temperature:-40℃~ +55℃
 Environment Relative Humidity: ≤95%
 Atmospheric Pressure: 500~1060hPa

Package:
 Each unit is packed one case
   Size:110cm× 78cm × 86cm
   Gross weight: ≤90Kg