Anesthesia Machine
■Main body: High-strength engineering plastic rack, light, beautiful and corrosion resistant
■Scope of application: adult, child
■Gas source: O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
■Flow meter: O2: 0.1~10L/min N2O: 0.1~10L/min
■O2, N2O linkage and N2O Stopper: When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off.
■Flow rate of rapid oxygen supply: 25~75L/min
■Low oxygen pressure alarm: There will be sound alarm when the oxygen pressure < 0.2MPa
■Vaporizer: It has the function of automatic compensation based on pressure, temperature, and flow rate.
The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane,
Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And
vaporizers that are imported with original packaging are also available.
■Respiratory circuit: working mode: all-close, semi-close, semi-open; releasing pressure: 1-4 kPa
■Respiratory bellows: bellows for adults, bellows for children; tidal volume range:0~1,500 ml
******************************************************************************************
Ventilator
■Display mode: High-definition 5.7” LCD screen display
■Ventilation mode: IPPV,SIPPV,SIMV,PEE,MANUAL,SIGH
■Ventilation parameters: Tidal volume: 50~1500ml
Rate: 2~99bmp
SIMV rate: 2~20bpm
I:E: 2:1~1:8
Inspiratory trigger pressure: -1.0~2.0 kPa
PEEP: 0~2.0 kPa
Pressure range: 1.0~6.0 kPa
SIGH: 1.5 times the inspiratory time
■Parameters for ventilation monitoring:
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of
airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure
(oxygen concentration)
■Security Alarm System:
Oxygen Concentration alarm: upper limit setting range (50%~100%)
low limit setting range (15%~50%)
Airway pressure alarm: upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm: upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm: it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm: it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
■ Power: AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
*Items within brackets are optional.