Anesthesia Machine
■Main body: High-strength engineering plastic rack, light, beautiful and corrosion resistant
■Scope of application: adult, child
■Gas source: O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
■Flow meter: O2: 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
■O2, N2O linkage and N2O Stopper: When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Mpa, the flow of nitrous oxide would be cut off.
■Flow rate of rapid oxygen supply: 25~75L/min
■Low oxygen pressure alarm: There will be sound alarm when the oxygen pressure < 0.2MPa
■Vaporizer: It has the function of automatic compensation based on pressure, temperature, and flow
rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,
Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by
the customer. And vaporizers that are imported with original packaging are alsoavailable.
■Respiratory circuit: working mode: all-close, semi-close, semi-open; releasing pressure: 0.5~7 kPa
■ Respiratory Bellows: bellows for adults, bellows for children; tidal volume range:0~1500 ml
***************************************************************************
Ventilator
■Display mode: High-definition TFT color LCD screen display
■Ventilation mode : IPPV, SIPPV, SIMV, PEEP, MANUAL,
SIGH, (IMV, VCV, PCV)
■ Ventilation parameters: Tidal volume: 50~1500 ml (20~1500 ml)
Rate: 1~99 bpm
SIMV rate: 2~20 bpm
I:E: 2:1~1:8 (4:1~1:8)
Inspiratory trigger pressure: -1.0~2.0Kpa
PEEP: 0~2.0Kpa
Pressure Range: 1.0~6.0Kpa
SIGH: 1.5 times the inspiratory time (70-150 adjustable)
Inspiratory platform: 0~50%
■Parameters for ventilation monitoring:
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway,
average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow
-volume oop,pressure-volume loop)
■Monitoring of oxygen concentration: 21%~100%
■Security Alarm System:
Oxygen Concentration alarm: upper limit setting range 21%~100%
low limit setting range 10%~80%
Airway pressure alarm: upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm: upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm: It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm: It will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
■ Power: AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
■(Mechanical Arms)
■(End-expiratory CO2 monitoring)
*Items within brackets are optional.