Anesthesia Machine
Main body: High-strength engineering plastic rack, light,beautiful and corrosion resistant
Scope of application: adult, child
Gas source: O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
Flow meter: O2: 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
O2,N2O linkage and N2O Stopper: When using nitrous oxide, oxygen concentration>25%;When the oxygen pressure<0.2Mpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply: 25~75L/min
Low oxygen pressure alarm: There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer: it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane,Isoflurane and Sevoflurane,two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
Respiratory circuit: working mode: all-close, semi-close, semi-open APL: 0.5~7 kPa
Respiratory Bellows: bellows for adults, bellows for children Tidal volume range:0~1500 ml
**************************************************************************************************************************
Ventilator
Display mode: High-definition 5.7” LCD screen display
Ventilation mode: IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
Ventilation parameters: Tidal volume: 50~1500 ml
Rate: 2~99 bpm
SIMV rate: 2~20 bpm
I:E: 2:1~1:8
Inspiratory trigger pressure: -1.0~2.0 kPa
PEEP: 0~2.0 kPa
Pressure range: 1.0~6.0 kPa
SIGH: 1.5 times the inspiratory time
Parameters for ventilation monitoring:
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of
airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure,
(oxygen concentration)
Security Alarm System:
Oxygen Concentration alarm: upper limit setting range (50%~100%)
Low limit setting range (15%~50%)
Airway pressure alarm: upper limit setting range 1.0~6.0 kPa
Low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm: upper limit setting range 3.0~30L/min
Low limit setting range 1.0~10L/min
Sustained high-pressure alarm: It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm: It will give sound and light alarm if there is no tidal volume input for 15 seconds
Power alarm
Power: AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
(Mechanical Arms)
(End-expiratory CO2 monitoring)
*Items within brackets are optional.